CBP1_v2 PEREL 10
• Never use too much water.
• Do not bend the cuff or sleeve. Wrap the inside to the outside.
• Do not dismantle the unit, remove the cuff or try to repair yourself. Contact your distributor.
• Do not operate the unit under severe climatic circumstances or at extreme temperatures,
humidity or direct sunlight.
18. Troubleshooting and Error Messages
Trouble/Error Possible Cause Solution
No indication when powered
on
Batteries are not properly
installed
• Drained batteries
Check the polarity of the
batteries
• Remove the batteries and
wait 10 seconds before
reinstalling
Err 0
No pulse or detected pulse
too low
Refit the cuff and repeat
the measurement
Err 1
Cuff is leaking air or
inflation is too slow
Replace the batteries or
contact your distributor
Err 2
Cannot measure systolic
Refit the cuff and repeat
the measurement
Err 3
Cuff pressure is too high
Refit the cuff and repeat
the measurement
Err 4
Cannot measure systolic
or is outside of range
Refit the cuff and repeat
the measurement
Err 5
Deflation is too fast
Contact your distributor
Err 6
Deflation is too fast
Contact your distributor
Err 7
Systolic is outside of range
Refit the cuff and repeat
the measurement
Err 8
Main pressure is outside of
range
Refit the cuff and repeat
the measurement