Velleman AVM360 Bedienungsanleitung Seite 7

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 30
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 6
AVM360 VELLEMAN 7
4. Read the hFE scale. The value of the reading is Ic / Ib, which is the DC amplification degree of the tested
transistor.
4.8 Diode test
1. Set the range selector at selected range position:
x 1K for 0 - 150 µA test,
x 100 for 0 - 1.5 mA test,
x 10 for 0 - 15 mA test,
x 1 for 0 - 150 mA test.
2. Connect the diode to the tester.
For I
F
(forward current) test, connect the "N" terminal of the tester to the positive polarity of the diode and the "P"
terminal to the negative polarity of the diode. For I
R
(reverse current) test, reverse the connection.
3. Read I
F
or I
R
one the LI scale provided.
4. Read the linear (forward) voltage of the diode on the LV scale while testing I
F
or I
R
.
5. Specifications
5.1 DC Voltage
Ranges: 0.25 - 0.5 - 2.5 - 10 - 50 - 250 - 1000 V
Accuracy at FSD: 3
Sensitivity: 20 k / V
Extension: 25 kV (with HV probe extra)
5.2 AC Voltage
Ranges: 10 - 50 - 250 - 1000 V
Accuracy at FSD: 4
Sensitivity: 9 k / V
Decibelmeter: -10 to +22 dB
0 dB = 1mW / 600
5.3 DC Current
Ranges: 50 µA (at 0.1 VDC position) - 2.5 mA - 25 mA - 0.25 A
Accuracy at FSD: 3
Volt Drop: 250 mV
5.4 Resistance
Ranges:
x 1 - 0.2 up to 2 k, midscale, at 20
x 10 - 2 up to 20 k, midscale, at 200
x 100 - 20 up to 200 k, midscale, at 2 k
x 1k - 200 up to 2 M, midscale, at 20 k
x 10k - 2 k up to 20 M, midscale, at 200 k
Accuracy at FSD: 3
Seitenansicht 6