DVM601
09/04/2009 ©2008 Velleman
®
Components nv
7
the red lead connection.
• Press the Hz/% button to switch to frequency; press again to see the
duty cycle (%).
Range Resolution
Range Resolution Range Resolution
400.0mV* 0.1mV 40.00V 10mV 600V 1V
4.000V 1.0mV 400.0V 100mV
*: no auto-ranging for AC
AC/DC current measurement
• Connect the red test lead to the µA/mA [6] input jack (max. 400mA) or
10A [7] input jack and the black test lead to the COM jack [5].
• Set the rotary switch [3] in the desired current range (<4mA: µA
,
<400mA: mA
, <10A: 10A ) and use the mode button [3] to
choose between AC and DC current.
• Open the circuit in which the current is to be measured and connect the
test leads to the circuit IN SERIES.
• Read the current value and the polarity of the red lead connection on the
LCD display.
Range Resolution
Range Resolution
400.0µA 0.1µA 400.0mA 100µA
4000µA 1µA 10A 10mA
Resistance, diode test and audible continuity test
• Connect the red test lead to the -jack [4] and the black test lead to the
COM jack [5] (the red lead has a positive polarity "+").
• Set the rotary switch [3] in the Ω range position. Do not apply input
voltage at this setting!
• Use the mode button [13] to choose between , diode or continuity test.
• If the resistance being measured is connected to a circuit, turn off the
power and discharge all capacitors before applying the test probes.
• Connect the test leads to the resistor to be measured and read the LCD
display.
Note:
wait a few seconds for a stable reading when measuring resistors >
1M.
Range Resolution
Range Resolution Range Resolution
400.0 0.1 40.00K 10 4.000M 10k
4.000k 1 400.0k 100 40.00M 1M
• For diode tests, connect the red test lead to the anode of the diode to be
tested and the black test lead to the cathode of the diode. The approx.
forward voltage drop of the diode will be displayed (typically 0.400 –
0.700V). If the connection is reversed, the display will show “OL”.
Kommentare zu diesen Handbüchern